Spectrophotometry and Beer's Law

How do chemists measure concentration with light, and what is Beer's law?

IntermediateAnalytical ChemistryLast reviewed 4 October 2026

What is it?

A coloured solution looks coloured because it absorbs some wavelengths of light and lets the rest through. The more concentrated the solution, the more light it absorbs. Spectrophotometry measures this absorption to find the concentration of a substance, quickly and with very small samples.

A spectrophotometer shines light of one chosen wavelength through the sample and measures how much comes out:

  • I0I_0 is the intensity of light entering the sample, and II the intensity leaving it.
  • Transmittance: T=II0T = \dfrac{I}{I_0} (between 0 and 1), or as a percentage, %T=T×100 %\%T = T \times 100\ \%.
  • Absorbance: A=−log⁡T=log⁡I0IA = -\log T = \log \dfrac{I_0}{I}

Absorbance and transmittance are ratios, so they have no units. A = 0 means no light is absorbed; A = 1 means only 10 % gets through; A = 2, only 1 %.

Key idea

Beer’s law (the Beer–Lambert law): absorbance is directly proportional to concentration.

A=εbcA = \varepsilon b c

where ε\varepsilon is the molar absorptivity (L mol⁻¹ cm⁻¹, a constant for that substance at that wavelength), bb is the path length of the cell (cm, usually 1.00 cm) and cc is the concentration (mol/L). Double the concentration, double the absorbance.

Why does it matter?

  • Medicine. Blood tests for glucose, cholesterol and haemoglobin use colour reactions read by a spectrophotometer.
  • Environment. Nitrate, phosphate and iron in drinking water and rivers are measured this way.
  • Biology and industry. DNA and protein concentrations, food colourings and the progress of reactions are all followed by absorbance.

How does it work?

1. The spectrophotometer

Light from a lamp passes through a monochromator (a grating), which selects one wavelength. That light passes through the sample in a cuvette of path length bb and reaches a detector, which reports TT or AA. The wavelength chosen is usually the one at which the substance absorbs most strongly, λmax\lambda_\text{max}: there the measurement is most sensitive, and small errors in the wavelength setting matter least.

2. Checking the units in Beer’s law

A=ε×b×cunits: (L mol−1 cm−1)×cm×(mol L−1)\begin{aligned} &A = \varepsilon \times b \times c \\[4pt] &\text{units: } \left(\text{L mol}^{-1}\,\text{cm}^{-1}\right) \\[4pt] &\qquad \times \text{cm} \times \left(\text{mol L}^{-1}\right) \end{aligned}

All the units cancel, so AA has no unit, as it should.

3. The calibration curve

In practice, ε\varepsilon is rarely taken from a table. Instead:

  1. Zero the instrument with a blank: the solvent and reagents, without the substance being measured. This removes absorption by everything else.
  2. Measure a series of standards of known concentration.
  3. Plot absorbance against concentration and draw the best-fit straight line (by least squares, using a calculator or spreadsheet).
  4. Measure the unknown in replicate (at least in triplicate), find the mean absorbance and its standard deviation, and read the concentration from the line.

The unknown should lie within the range of the standards. If its absorbance is too high, dilute it by a known factor and measure again.

Think of it like this

Think of sunglasses. One pair of lenses dims the light by a certain fraction; two pairs stacked dim it again by the same fraction. A thicker or darker filter (a longer path length or a higher concentration) lets through less light. Absorbance is the scale that turns this “fraction of a fraction” behaviour into a simple straight line.

More precisely

Beer’s law holds best for dilute solutions and absorbances up to about 1. At high concentrations, molecules interact with one another, and at very high absorbances so little light reaches the detector that stray light causes errors; the calibration line then curves. Beer’s law also assumes monochromatic light, which is why the monochromator matters. The slope of the calibration line is the sensitivity, and the intercept should be close to zero if the blank was correct.

Visualise it

A spectrophotometer drawn left to right: a lamp, a monochromator containing a grating, a sample in a cuvette with path length b, and a detector reading A = 0.412. The beam entering the sample is labelled I0 and is thick; the beam leaving is labelled I and is thinner. Formulas: T = I divided by I0, and A = −log T = log of I0 divided by I.
The sample absorbs part of the light: I is smaller than I₀.
A calibration curve of absorbance against concentration in mg/L. Five standards at 1, 2, 3, 4 and 5 mg/L lie on a straight line, A = (0.1986 L/mg) c + 0.0020. The unknown, the mean of 3 replicates, has absorbance 0.412; a dashed line across to the calibration line and down to the axis gives c = 2.07 mg/L.
Standards give the line; the unknown's mean absorbance is read across and down.

Worked example

Worked example: From transmittance to absorbance

Question: A solution transmits 25.0 % of the light. What is its absorbance?

  1. T=25.0 %100 %=0.250T = \dfrac{25.0\ \%}{100\ \%} = 0.250
  2. A=−log⁡(0.250)=A = -\log(0.250) = 0.602 (no unit)

Worked example: Using Beer's law

Question: A dye has ε=1.20×104\varepsilon = 1.20 \times 10^{4} L mol⁻¹ cm⁻¹ at its λmax\lambda_\text{max}. A solution of the dye in a 1.00 cm cell has an absorbance of 0.540. Find its concentration.

  1. Rearrange: c=Aεbc = \dfrac{A}{\varepsilon b}

  2. Substitute:

    εb=(1.20×104 Lmol cm)×(1.00 cm)=1.20×104 L/molc=0.5401.20×104 L/mol=4.50×10−5 mol/L\begin{aligned} &\varepsilon b = \left(1.20 \times 10^{4}\ \tfrac{\text{L}}{\text{mol cm}}\right) \\[4pt] &\qquad \times (1.00\ \text{cm}) \\[4pt] &\quad = 1.20 \times 10^{4}\ \text{L/mol} \\[4pt] &c = \frac{0.540}{1.20 \times 10^{4}\ \text{L/mol}} \\[4pt] &\quad = 4.50 \times 10^{-5}\ \text{mol/L} \end{aligned}
  3. The cm cancels, and 1L/mol=mol/L\dfrac{1}{\text{L/mol}} = \text{mol/L}: 4.50 × 10⁻⁵ mol/L.

Worked example: An unknown from a calibration curve, measured in triplicate

Question: Standards of an iron complex (1.00 to 5.00 mg/L) give the best-fit line A=(0.1986 L/mg) c+0.0020A = (0.1986\ \text{L/mg})\,c + 0.0020. A water sample is measured three times: A = 0.412, 0.416 and 0.409. Find the iron concentration, and comment on the precision.

  1. Mean absorbance: Aˉ=0.412+0.416+0.4093=0.4123\bar{A} = \dfrac{0.412 + 0.416 + 0.409}{3} = 0.4123
  2. Standard deviation (sample, n−1n - 1): s=0.0035s = 0.0035; relative standard deviation =0.00350.4123×100 %=0.85 %= \dfrac{0.0035}{0.4123} \times 100\ \% = 0.85\ \%
  3. Rearrange the line: c=Aˉ−0.00200.1986 L/mgc = \dfrac{\bar{A} - 0.0020}{0.1986\ \text{L/mg}}
  4. Substitute: c=0.4123−0.00200.1986 L/mg=c = \dfrac{0.4123 - 0.0020}{0.1986\ \text{L/mg}} = 2.07 mg/L
  5. Report: Aˉ=0.4123±0.0035\bar{A} = 0.4123 \pm 0.0035 (n=3n = 3). An RSD below 1 % shows good precision; the result lies inside the range of the standards (1.00 to 5.00 mg/L), so the reading is reliable.

Common mistake

Common mistake: Using %T directly in the absorbance formula

Convert a percentage to a fraction first: for 25.0 %T, A=−log⁡(0.250)=0.602A = -\log(0.250) = 0.602, not −log⁡(25.0)=−1.40-\log(25.0) = -1.40. An absorbance is never negative for a real sample.

Common mistake: Thinking transmittance is proportional to concentration

Only absorbance is proportional to concentration. Doubling the concentration doubles A, but squares T (for example, T = 0.50 becomes T = 0.25).

Common mistake: Reading far outside the calibration range

A result outside the standards relies on the line continuing straight, which may not be true at high absorbance. Dilute the sample by a known factor (and multiply the result by it), or add more standards.

Notation note

  • Molar absorptivity ε\varepsilon is also called the molar extinction coefficient; its units can be written L mol⁻¹ cm⁻¹ or M⁻¹ cm⁻¹.
  • Some books write A=abcA = abc with aa the absorptivity in mass units (for example L g⁻¹ cm⁻¹), when cc is in g/L.
  • The blank is sometimes called the reference solution.

Remember this

Remember this

  • T=I/I0T = I/I_0; A=−log⁡TA = -\log T. Both have no units.
  • Beer’s law: A=εbcA = \varepsilon b c, with ε\varepsilon in L mol⁻¹ cm⁻¹, bb in cm, cc in mol/L.
  • Measure at λmax\lambda_\text{max}, zero with a blank, and use a calibration curve of standards.
  • Measure unknowns in replicate (at least triplicate) and report the mean ± standard deviation; stay inside the calibration range.

Test yourself

Check your understanding before moving on.

Flashcards

Spectrophotometry and Beer's Law: Flashcards

10 cards

  1. Question
    Define transmittance.
    Answer

    T = I ÷ I₀, the fraction of light that passes through the sample (no unit). %T = T × 100 %.

  2. Question
    How is absorbance related to transmittance?
    Answer

    A = −log T = log(I₀ ÷ I). It has no unit.

  3. Question
    State Beer's law.
    Answer

    A = εbc: absorbance is proportional to concentration (ε in L mol⁻¹ cm⁻¹, b in cm, c in mol/L).

  4. Question
    What is ε?
    Answer

    The molar absorptivity: a constant for a substance at a given wavelength, in L mol⁻¹ cm⁻¹.

  5. Question
    Why measure at λmax?
    Answer

    Absorption is strongest there, so the method is most sensitive, and small wavelength errors matter least.

  6. Question
    What is a blank, and why is it used?
    Answer

    The solvent and reagents without the analyte. It zeroes the instrument so only the analyte absorption is measured.

  7. Question
    What is a calibration curve?
    Answer

    A plot of absorbance against concentration for standards of known concentration; the unknown is read from its best-fit line.

  8. Question
    Why measure the unknown in replicate?
    Answer

    To assess precision: report the mean ± standard deviation (at least triplicate) instead of a single reading.

  9. Question
    What is the absorbance when 10 % of the light is transmitted?
    Answer

    A = −log(0.100) = 1.000.

  10. Question
    When does Beer's law break down?
    Answer

    At high concentrations or high absorbances (above about 1), and with non-monochromatic light: the calibration line curves.

Quiz

Spectrophotometry and Beer's Law: Quiz

7 questions

  1. Question 1EasyA solution transmits 50.0 % of the light. What is its absorbance?
    Show answer

    Answer: 0.301

    T = 0.500, so A = −log(0.500) = 0.301. Using 50.0 instead of 0.500 gives −1.70, which is impossible for a real sample.

  2. Question 2EasyIf the concentration of a solution is doubled (same cell), the absorbance:
    Show answer

    Answer: doubles

    Beer's law, A = εbc: A is directly proportional to c.

  3. Question 3MediumWhat are the units of molar absorptivity, ε?
    Show answer

    Answer: L mol⁻¹ cm⁻¹

    A has no unit, so ε = A ÷ (b × c) has units 1 ÷ (cm × mol/L) = L mol⁻¹ cm⁻¹.

  4. Question 4MediumA = 0.450, ε = 1.50 × 10⁴ L mol⁻¹ cm⁻¹, b = 1.00 cm. What is c?
    Show answer

    Answer: 3.00 × 10⁻⁵ mol/L

    c = A ÷ (εb) = 0.450 ÷ (1.50 × 10⁴ L mol⁻¹ cm⁻¹ × 1.00 cm) = 3.00 × 10⁻⁵ mol/L.

  5. Question 5EasyWhat is the purpose of the blank?
    Show answer

    Answer: to zero the instrument so that only the analyte absorbance is measured

    The blank contains everything except the analyte, so absorption by the solvent, reagents and cell is subtracted.

  6. Question 6MediumThe same solution is moved from a 1.00 cm cell to a 2.00 cm cell. Its absorbance changes from 0.250 to:
    Show answer

    Answer: 0.500

    A is proportional to the path length b: doubling b doubles A.

  7. Question 7HardAn unknown gives A = 1.85, but the standards only go up to A = 0.95. What should you do?
    Show answer

    Answer: dilute the unknown by a known factor, measure again and multiply by the factor

    The reading must fall inside the calibration range; Beer's law may not hold at high absorbance. A longer cell would raise A even further.

Notes and downloads

  • Worksheet

    Spectrophotometry and Beer's Law Worksheet

    9 questions on transmittance and absorbance, Beer's law, dilution, calibration curves with replicates, and a copper analysis. Answer key included.

    IntermediateFree

References

  1. Brown, T. L.; LeMay, H. E., Jr.; Bursten, B. E.; Murphy, C. J.; Woodward, P. M.; Stoltzfus, M. W. Chemistry: The Central Science, 15th ed.; Pearson, 2022.

Spotted a mistake? Let us know and we'll fix it.